ESCORT Prism™ & Prism SEPatient MonitorReference ManualPrism SE Model20411204122041320414Prism Model2040120403

Seite 2

ixChapter 9 – Clinical Information System... 9-1Introductio

Seite 3

7-3SPO2 MessagesSPO2 MESSAGE MEANINGSPO2 SRCH(Nellcor only)The monitor cannot locate the patient’s pulse. The patient’sperfusion may be too poor to de

Seite 4

7-4Noninvasive Blood Pressure (NIBP) MessagesNIBP MESSAGE MEANINGET = MM:SS Indicates elapsed time (ET) since last NIBP measurement.AET = MM:SS Indica

Seite 5

7-5SureTemp MessagesSURETEMP MESSAGE MEANINGT1 INITIAL This message display for a few moments as the probe warms up. T1 WAIT This message displays mom

Seite 6

7-6Cardiac Output (CO) MessagesCO MESSAGE MEANINGNO CATH No catheter connected to the monitor.NO PROBE No temperature probe (bath or injectate) connec

Seite 7

8-1Chapter 8 – Defibrillator/PacerIntroductionThe portable ESCORT defibrillator/pacer can attach to any ESCORT 300, II 300, or Prismmonitor to provide

Seite 8

8-2connection, the unit must be operated by battery power.CAUTION: Make sure to use an ECG cable with an inline resistor sothe monitor is protected f

Seite 9

8-3Defibrillator/Pacer Front Overview1. Defibrillator/Pacer Carrying Handle2. APEX Defibrillator Paddle3. Energy CHARGE (transparent/lighted) and D

Seite 10

8-4Defibrillator/Pacer Rear Overview1. AC Receptacle2. Defibrillator Test Lamp3. Voltage Select/Fuse Holder4. Battery5. Battery Lock

Seite 11

8-5Defibrillator/Pacer Front Panel1. External Pace Rate Adjustment Knob2. QRS Indicator3. External Pacing Mode Key4. Pace Current Adjustment Knob5

Seite 12

8-617. Defib Energy CHARGE KeyUsing Defibrillator and Pacer Paddles and ElectrodesExternal PaddlesThe STERNUM defibrillating paddle contains a dischar

Seite 13

xTHIS PAGE LEFT BLANK INTENTIONALLY

Seite 14

8-7To Apply Disposable Adhesive Electrodes to the Patient:1. In a non-emergency application, clean and dry skin in electrode areas, preferably with a

Seite 15

8-8Operating the Defibrillator1. Ensure that the paddle cable is connected to the back of the defib/pacer behind the monitor.2. Apply conductive gel

Seite 16

8-96. Place defibrillator paddles firmly on the patient’s chest with the STERNUM paddle near theupper sternum and slightly toward the patient’s right

Seite 17

8-103. Number “2”—select the energy to be delivered using the ENERGY SELECT knob on thefront panel.NOTE: If the energy select knob is changed after

Seite 18

8-117. Observe the patient and check Prism monitor to determine results. If repeat defibrillation isnecessary, readjust the energy level as prescribe

Seite 19

8-126. Follow defibrillation procedure in the section “Operating the Defibrillator”.NOTE: The defib/pacer automatically goes to defib (async) modeaf

Seite 20

8-134. Select the desired pace rate with the external pace rate switch on the defib/pacer front panel.5. Select either asynchronous (continuous) or

Seite 21

8-14• the pacing cable is properly connected to the electrodes and to the pace outputconnectorPacing and Defibrillating with the Same ElectrodesIn or

Seite 23

9-1Chapter 9 – Clinical Information SystemIntroductionThe CIS Communication Module when attached to a Prism or Prism SE Patient Monitor, allowsfor the

Seite 24

1-1Chapter 1 – Physical DescriptionThis section describes the controls, indicators, and features located on the front, back, and sidepanels of the Pri

Seite 25

9-21. CIS Status Message Zone2. CIS Parameter Label3. Softkeys for Interfacing with the CIS Menu4. CIS Menu Display ZoneNOTE: At power up, the CI

Seite 26

9-33. Press the PAGE HOME key to remove the CIS Menu display and replace theregularly monitored parameters in the waveform display zone. The displa

Seite 27

9-4The action taken upon the selection of a softkey is determined by the current menu selectionsdisplayed on the CIS menu located in the Trace 2 wavef

Seite 28

9-5A brief overview of the patient confirmation process is included here to provide a betterunderstanding of the significance and overall scope of thi

Seite 29

9-6location, must be queried using the “Query By Patient Identification” method describedbelow.Monitors that are situated between multiple room/bed lo

Seite 30

9-7Confirming the Patient:CIS STATUS MESSAGE: CIS STATUS MESSAGE:“CONFIRM PT” (flashing) Displays Confirmed Patient’s Name and IdentificationOnce the

Seite 31

9-8Using the VSC Function to Set Up CIS Charting IntervalsEach monitor includes the ability to store vital signs in a tabular format at user-selectabl

Seite 32

9-9In the sample VSC table illustration, notice that there is an unlabeled column to the left of theTIME column. This column is used to display a lett

Seite 33

9-101. Press the SYSTEM SETUP key then select the CLEAR PT DATA softkey. Be sureto select “YES” when faced with the “Are You Sure?” question.CLEARPTD

Seite 34

9-11NOTE: Episodic measurements will not post to a VS chart morethan once for all automatic entries, even if they are displayed andwithin the Valid T

Seite 35

1-210. Patient Type Selection Key (Adult, Ped, Neo)11. Start/Stop Manual Recordings12. Display System Setup Menu13. Return to Home Page14. Seven Param

Seite 36

9-125. Press the PAGE HOME key when you have finished setting up VSC toreturn to the main monitoring screen.Assigning VSC to a Waveform Trace1. Pr

Seite 37

9-13Refer to section “Deleting Vital Signs Entries”, for details on the DEL VSC ENTRYoption.Manually Entering Vital Signs Data Using the CIS MenuVital

Seite 38

9-142. Press the CIS softkey associated with the ENTER VITAL SIGNS function. Thefollowing CIS menu will be displayed with the corresponding CIS softk

Seite 39

9-151. Display the VSC parameter in the monitor waveform zone (refer to section “Assigning VSCto a Waveform Trace”).2. Press the softkey adjacent to

Seite 40

9-16NOTE: If the network connection is lost per the above situation, theCIS Status Message at the top of the monitor will no longer simplydisplay the

Seite 41

9-17Using the Temporary Name FeatureCIS STATUS MESSAGE:Displays Temporary Patient’s Name (Flashing)Patient vital signs information that is taken on an

Seite 42

9-18The optional keyboard can be used to facilitate entry of the temporary name. To use the keyboard:1. Access the TEMP NAME function as described

Seite 43

9-19To change the filing mode for this record, select the F2 key. The filing mode will togglebetween FINAL and REVIEW modes. Refer to section “Selecti

Seite 44

9-20 HR HIGH LIMIT is 1 byte long and represents heart rate HIGH alarm limit (not currentlysupported at this time). HR LOW LIMIT is 1 byte long an

Seite 45

9-21Sample Waveform PacketPacket ID2 bytesWaveformNumber1 byteLength1 byteDataCheckSum1 byte1b 57 00 13 83 83 81 80 7d 98 b3 cb b3 96 7a 7e 80 81 83 8

Seite 46

1-3ALARM SETUPDisplays the alarm status for all vital sign parameters. Alarm ON/OFF status, alarm prioritylevel, high and low limits, and recording ty

Seite 47

9-22Table 5: Remote StatusRemote StatusHex Value DescriptionNull Status 00h No activityLow Battery 10h Battery voltage is lowAlarm Suspend 70h Alarm

Seite 48

9-23Additional WaveformsAs previously stated, the CIS module transmits two type of packet information, a parameterpacket and a waveform packet. The EC

Seite 50

10-1Chapter 10 – TroubleshootingGeneral SolutionsProblem Possible Causes Action/SolutionMonitor does not turn on.AC power may not beconnected.Verify A

Seite 51

10-2Respiration SolutionsProblem Possible Causes Action/SolutionErratic Respiration rate andwaveform.Poor electrode placement orconductivity.Confirm p

Seite 52

10-3NO SENSOR is displayed. Sensor not connected.Damaged Sensor or cable.Unknown sensor connected.Check sensor connections.Verify red light on sensori

Seite 53

10-4Mainstream ETCO2 Solutions (Prism Only)Problem Possible Causes Action/SolutionNO SENSOR message. Sensor not connected. Confirm proper sensorconnec

Seite 54

10-5ETCO2 Solutions (Prism SE Only)Problem Possible Causes Action/SolutionNo Sensor message. No filterline connected. Connect a filterline to monitor.

Seite 55

10-6Cardiac Output (CO) SolutionsProblem Possible Causes Action/SolutionNO CATH message. Catheter not connected. Confirm catheter connectedproperly.NO

Seite 56

10-7power.Defib/Pacer SolutionsProblem Possible Causes Action/SolutionDefibrillator will not charge(beeps intermittently) whenCHARGE key is pressed.De

Seite 57

1-4RECORDInitiates a 16-second recording of any one or two parameter waveforms. To stop a recording inprocess, press RECORD prior to completion of a m

Seite 58

10-8THIS PAGE LEFT BLANK INTENTIONALLY

Seite 59

11-1Chapter 11 – Supplies, Accessories, & OptionsThe supplies, accessories, and options listed in this section can be ordered by callingMedical Da

Seite 60

11-2PartNumberDefibrillator / Pacer Accessories / SuppliesE2700-03GCX Defib Slide Mount Adapter Plate – Allows monitor equipped withDefib base to atta

Seite 61

11-3PartNumberECG Accessories – CE ApprovedE2900-04 5-Lead ECG Safety Cable (Red, Yellow, Green, Black, White)E2900-03 3-Lead ECG Safety Cable (Red, Y

Seite 62

11-4PartNumberMicrostream ETCO2 Accessories / Supplies (Prism SE Only)E2910-20Adult / Pediatric Filterline & Airway Adapter Set – for intubated CO

Seite 63

11-5PartNumberNIBP Accessories / Supplies – Screw Type FittingsE2800-06 Adult / Pediatric Hose Assembly (12 foot)E2800-11 Neonatal Hose Assembly (8 fo

Seite 64

11-6PartNumberSPO2 Accessories / Supplies – Masimo E2810-50 Adult Finger Sensor-ReusableE2810-51 Adult Disposable Adhesive Sensors (box of 20)E2810-5

Seite 65

11-7PartNumberVideo CableE2800-SC2525 Foot Standard CableE2800-SC5050 Foot Standard CableE2800-PC5050 Foot Plenum CablePartNumberRemote Display Power

Seite 66

11-8PartNumberOptions (Prism Only)OPT25 BEDRAIL HANDLE – Need 9040-60 upgrade on older unitsOPT37CRG and OXYCRGS – Provides CRG (Cardiorespirogram) to

Seite 67

11-920001-MMULTIPARAMETER MODULE – MMPM for use with Masimo SpO2 or when no SpO2 is ordered.Includes single or dual lead ECG and Respiration.20001-NMU

Seite 68

1-5IndicatorsThe monitor includes up to four indicators, which illuminate to notify the operator of a specificactivity or situation. These indicators

Seite 69

11-10PartNumberOptions (Prism SE Only)OPT53-SEVGA Output for Prism SE series monitors – Permits all Prism SEwaveforms and numeric data to be displayed

Seite 70

11-11PartNumberWireless Communication Modules 20011VVISION TRANSCEIVER MODULE – (900 MHz BAND) SpreadSpectrum communication with Vision central statio

Seite 71

11-12THIS PAGE LEFT BLANK INTENTIONALLY

Seite 72

12-1Chapter 12 – SpecificationsPRISM & PRISM SE SPECIFICATIONSPHYSICAL (Prism Only)Size: 9.3 x 11.8 x 10.2 in./23.6 x 30.0 x 25.9 cm (HWD)Weight:

Seite 73

12-2Range: 10 to 300 BPMAccuracy: ±2% or ±2 BPMSweep Speed: 12.5, 25 mm/sec selectableDisplay Update Information: Display update interval is 2 sec

Seite 74

12-3Range: 20 to 50° C (68 to 122° F)Alarm Limit Range 20-44° CAccuracy: ±0.1° CProbe: YSI 700 or 400, autosensingConnector: 1/4 inch (0.6 cm) stereo

Seite 75

12-4Pulse Tone: Pitch varies with SpO2 valueECG Sync: NELLCOR C-LOCKSpO2 (Nellcor MP405)Display: % SpO2, Plethysmographic Waveform or Pulse Amplitu

Seite 76

12-5PED 0-150 mmHgNEO 0-150 mmHgMean: ADULT 0-200 mmHgPED 0-170 mmHgNEO 0-170 mmHgDisplay Scale: 15, 30, 60, 120, 180, 240 mmHg selectableInput Conne

Seite 77

12-6Operating Pressure: 430 to 795 mmHgMaximum Change Rate in Pressure: 12.4 mmHg/minOperating Altitude: -1,250 to 15,000 ft.Maximum Change Rate in

Seite 78

12-7Weight: 0.85 lb (0.39 kg) with full roll of paper (Prism SE Only)AutoNet TRANSCEIVER MODULEIndicator: Green light indicates comm. StatusFrequenc

Seite 79

1-6Prism & Prism SE Display Screen1. ECG Lead in Use2. Softkey Labels for Parameters Displayed in the Waveform Display Zone (WDZ)3. Alarm Statu

Seite 80

12-8POWERModular Batteries: 1-each 12V; 2.0 AH Rechargeable Sealed Lead AcidModular Battery Capacity: 35 charges @ 360 joules or 5 hours pacing at10

Seite 82

1-71. ECG Lead in UseIndicates the ECG lead displayed in the top trace of the WDZ. Choices include lead I, II, III, orV LD (and depending on hardware

Seite 83

1-87. System Message AreaA message that is not specific to a single parameter, such as “ALARMS SUSPENDED 180”, isdisplayed in a reverse-video box, ju

Seite 84

ESCORT Prism™ & Prism SEPatient MonitorReference ManualFederal law restricts this device to sale by or on the order of a physician.E9030-33R Rev F

Seite 85

1-9Prism SE Rear Panel1. External DC` Input Connector (optional)2. Fuse Holder3. AC Power Cord Input Connector4. Battery Compartment5. Model &

Seite 86

1-101. External DC Input (12 - 28V) ConnectorAllows an external DC power source to be connected for extended use during transport where ACpower is no

Seite 87

1-114. Battery CompartmentThis compartment houses one battery (or up to two batteries on earlier units) which providepower in transport situations or

Seite 88

1-12Prism Rear Panel1. External DC Input Connector (optional)2. Equipotential Connector (grounding terminal) (optional)3. Batteries4. Defibrillato

Seite 89

1-131. External DC Input (12 - 28V) ConnectorAllows an external DC power source to be connected for extended use during transport where ACpower is no

Seite 90

1-147. Auxiliary Output ConnectorProvides analog waveform or alarm relay closure interface signals from the monitor to otherclinical instruments.WARN

Seite 91

1-15Prism SE Side Panel

Seite 92

1-16Connector PanelThe Prism SE can monitor several parameters. They include ECG, SPO2, IBP (Invasive BloodPressure), NIBP (Noninvasive Blood Pressure

Seite 94

1-18Multiparameter ModuleThe Multiparameter Module is configured for several parameters; however your MultiparameterModule may be configured for fewer

Seite 95

iiPrefaceThe ESCORT Prism and Prism SE Series monitors are intended only as an adjunct to patientassessment. They cannot replace skilled nursing care

Seite 96

1-19Prism’s Modular ComponentsThere are several modular (user-movable) components that can be shared by all Prism monitors.• Multiparameter Module (M

Seite 97

1-20Prism SE’s Modular ComponentsThere are several modular (user-movable) components that can be shared by all Prism & PrismSE monitors.• Transce

Seite 99

1-22THIS PAGE LEFT BLANK INTENTIONALLY

Seite 100

2-1Chapter 2 – General Instructions for Using the Patient MonitorDetermine the Power SourceThe Prism and Prism SE can be powered in one of three ways:

Seite 101

2-2 To install the Multiparameter Module:1. Align the Multiparameter Module’s front panel with the parameter labels facing you andright-side up. In

Seite 102

2-3ALARM ON/OFF—Allows alarm to be turned ON/OFF.ALARM LIMITS—Allows adjustment of limits through use of ↑↓ arrow keys.SIZE/SCALE—Allows adjustment of

Seite 103

2-42. Press the TRACE SEL softkey to open the following menu.DISPVSCDISPOCRGTRACE2AUTOTRACE3AUTOTRACE4AUTOMORE3. Press the softkey under desired TRA

Seite 104

2-56. Use the UP ARROW ↑ ↑ and DOWN ARROW ↓ ↓ softkeys to adjust the date and timesettings.7. Press the PAGE HOME key.Monitor Default SettingsIf you

Seite 105

2-6 To adjust tone volumes for all parameters:1. Press the SYSTEM SETUP key to open the System Setup menu.

Seite 106

iiiSafety Considerations (Prism & Prism SE)Read the following sections before using the Prism or Prism SE monitors. These cautions andwarnings are

Seite 107

2-7CLEARPTDATASTOREVSDISPVSCPTSETUPHRAUTOECGTRACESELMORE2. Press the MORE softkey to open the following menu.ADULTPEDNEOVOLDISPOCRGRECSETUPTIMER CLOC

Seite 109

3-1Chapter 3 – Vital Sign Parameter UsageNOTE: The Prism SE monitor can be used during defibrillation, butcertain parameter readings may be inaccurat

Seite 110

3-2Patient PreparationWhen doing any patient monitoring, proper skin prep and electrode placement are critical toensuring patient safety as well as ac

Seite 111

3-3Neonatal Electrode Placement To monitor a patient’s ECG:1. Open the ECG SETUP menu by pressing the softkey next to the ECG label.ALMONOFFALMLIMS

Seite 112

3-4Selecting a Telemetry Patient to Monitor“Telemetry-monitored” patients must be prepared properly for telemetry functions to work. A3-lead or 5-lead

Seite 113

3-5RespirationRespiratory monitoring may be done either by impedance pneumography via ECG leads of RA(right arm) and LA (left arm) or through ETCO2. L

Seite 114

3-6Pulse Oximetry (SPO2)A noninvasive method for continuously measuring oxygen saturation and pulse rate. Hospitalpolicies, along with the sensor&apos

Seite 115

3-7MODE for Nellcor MP405 Averaging Time Application*NORM (NOT USED) (NOT USED) (NOT USED)FAST (NELLCOR Mode 2) Typically 7-8 seconds foraverage pulse

Seite 116

3-8SPO2 Sensor Accessories—NellcorNellcor Sensors – Contact NellcorSensor Model Patient sizeOxisensor® II oxygen transducers(Sterile, Single-use only)

Seite 117

ivCAUTION: Always locate the Prism or Prism SE and its power cordaway from any electrosurgery and diathermy equipment and its powercord and cables.CA

Seite 118

3-9 There is arterial occlusion proximal to sensor Patient is in cardiac arrest or is in shockSPO2 Sensor Accessories—MasimoName/Type Model Number

Seite 119

3-10Temperature MonitoringThe Prism and Prism SE can accommodate a variety of continuous temperature probes, as well asMDE SureTemp oral/axillary and

Seite 120

3-11T1 WAIT. DO NOT place probe into position until message reads T1 READY and anaudible tone is heard.3. When ready, place probe in position. ORAL a

Seite 121

3-12Each NIBP reading, recording, and trend printout will display an ET (elapsed time) number. Thisnumber indicates the number of minutes that have p

Seite 122

3-13Selecting an Appropriate NIBP CuffFor patient safety and accurate NIBP readings, it is essential to select an appropriate cuff forpatient type and

Seite 123

3-14Displaying the NIBP TableThe Prism and Prism SE monitors stores up to 100 NIBP readings, as well as any otherparameters being monitored, with thei

Seite 124

3-1510. Press the SITE softkey to cycle through site labels for invasive pressure line. Highlightdesired site.11. Press the MORE softkey to access sec

Seite 125

3-164. Press the CURSOR  softkeys to align cursor bar with end expiration. Note value displayedon cursor bar and at bottom of screen changes as cur

Seite 126

3-17If a BP in TRACE 4 or 5 is turned OFF, that BP will be removed from the waveform zone.The remaining BP will be expanded into the full Trace 4 area

Seite 127

3-18Injections are labeled as RUN 1, 2, etc. Subsequent injections are completed once START isredisplayed. A total of 6 RUNS can be displayed. Further

Seite 128

vTable of ContentsChapter 1 – Physical Description...

Seite 129

3-19End Tidal Carbon Dioxide (ETCO2) (Prism Only)End tidal carbon dioxide (ETCO2) monitoring can show early indications of circulatory andpulmonary pr

Seite 130

3-20WARNING: Do not allow CO2 sensor to come in contact withpatient’s skin for prolonged period of time. CO2 sensor heats toprevent fogging of airway

Seite 131

3-211. Inspect circular window on airway adapter for cracks or blemishes. Replace if necessary.2. Attach airway adapter and sensor in line proximall

Seite 132

3-22ETCO2 (CO2) Sensor and Airway Adapter Maintenance (Prism Only)CO2 sensor is a precision instrument, which requires proper maintenance to verify ac

Seite 133

3-23UNIT%mmHgO2ONOFFN2OONOFFETCO2 allows numeric information and waveform to be displayed. ETCO2 waveform can beplaced in a trace slot. Refer to confi

Seite 134

3-241. Position nasal cannula on patient.2. Connect ETCO2 tubing to Prism SE’s connector panel.3. Observe WARM UP message on screen. ETCO2 monitori

Seite 135

3-25Vital Sign Chart (VSC)VSC is a feature that allows storage, display, documentation and editing of vital signs. Themonitor stores 100 sets of tabul

Seite 136

3-26(bell) Entry made when an alarm is violated (alarming vital sign is shown in inverse video).n Entry made when a user-initiated NIBP reading is com

Seite 137

3-27CLEARPTDATASTOREVSDISPVSCPTSETUPHRAUTOECGTRACESELMORE2. Press the CLEAR PT DATA softkey and press YES softkey. 3. Press the PAGE HOME key. Oxyc

Seite 138

3-285. To remove temporary display, press the PAGE HOME key.  To display OxyCRG in Trace 2:1. Monitor ECG, Respiration, and SpO2.2. Press the SYS

Seite 139

viAC Power Cord Input Connector and Optional Power Cord Mounting Bracket... 1-13Auxiliary Output Connector ...

Seite 140

3-29AlarmsAlarm ON/OFF status and limits can be set from individual parameter menus or from ALARMSETUP page. Do not temporarily silence or disable aud

Seite 141

3-30understand the ramifications of turning alarms off, silencing alarms, suspending alarms, andresetting alarm criteria from factory defaults. Please

Seite 142

3-31(FLASHING)A solid bell will flash if one or more alarm criteria are violated for associatedparameter, and an alarm message will display in paramet

Seite 143

3-32 Continuous recordings:1. For ECG, press and hold the RECORD key for approximately three seconds.2. For all other parameters, press the RECORD

Seite 144

3-33CLEARPTDATASTOREVSDISPVSCPTSETUPHRAUTOECGTRACESELMORE2. Press the MORE softkey to display the following menu.ADULTPEDNEOVOLDISPOCRGRECSETUPTIMER

Seite 145

3-341. Press the door release to open recorder. 2. Remove the empty spool. 3. Install new roll with paper unrolling from bottom side. 4. Pull out

Seite 146

3-35THIS PAGE LEFT BLANK INTENTIONALLY

Seite 147

4-1Chapter 4 – Additional User InformationWireless Communication to a Central StationThe Prism and Prism SE monitors have the capability to send vital

Seite 148

4-2Maintenance and CalibrationThere are no user-serviceable parts in Prism or Prism SE monitors. MDE recommends a yearlyperformance check to verify al

Seite 150

viiMeasuring a Patient’s Temperature with Continuous Probes... 3-10Measuring a Patient’s Temperature with S

Seite 151

5-1Chapter 5 – Main Setup MenusThis section describes the main system setup and vital sign menus on the Prism and Prism SEmonitors. The menus are the

Seite 152

5-2ECG Setup MenusECG function menus are accessed by pressing the softkey adjacent to the ECG label. The menusare as follows:ALMONOFFALMLIMSIZELEADSEL

Seite 153

5-3RR AUTO LEADS—Selects RESP source, ECG leads, CO2, or AUTO detect.APNEA DELAY 20S—Selects allowable apnea period before alarming.MORE—Displays the

Seite 154

5-4SITE OFF ART—Adds label next to numeric to identify IBP sites.ZERO ALL—Zeroes all connected transducers.ZERO BP1—Zeroes the transducer offset.MORE—

Seite 155

5-5NIBP ON OFF—Press the NIBP softkey to turn ON or OFF.Continuous Temperature Setup MenusContinuous Temperature menus may be accessed by pressing the

Seite 156

5-6ALMONOFFALMLIMSCALE5075MM/S12.56.25CO2ONOFFMOREALM ON OFF—Toggles CO2 alarms ON or OFF.ALM LIM—Accesses alarm limits menu.SCALE 50/75—Selects CO2 w

Seite 157

5-7CONST .470 (ARROW ON 4)—Use to change the tenths column of the constant.CONST .470 (ARROW ON 7)—Use to change the hundredths column of the constant

Seite 158

5-8THIS PAGE LEFT BLANK INTENTIONALLY

Seite 159

6-1Chapter 6 – Factory DefaultsWhen the Prism & Prism SE monitor is powered on for the first time, it will be in the Neonatalmode, and all paramet

Seite 160

6-2Recorder Factory DefaultsRECORDER SETUP ADULT PED NEO DESCRIPTIONRECD DELAY ON ON ON Four second delayALRM RECDNG ON ON ON Alarm recording enable/d

Seite 161

viiiSystem Setup Factory Defaults ... 6-1Recorder Fac

Seite 162

6-3ECG Setup Factory DefaultsECG SETUP ADULT PED NEO DESCRIPTIONALARMS ON ON ON Enables ECG alarmsALARM LEVEL 2 2 2 Sets alarm priority level (1, 2, o

Seite 163

6-4SPO2 Setup Factory DefaultsSPO2 SETUP ADULT PED NEO DESCRIPTIONALARMS OFF ON ON Enables SPO2 alarmALARM LEVEL 2 2 2 Sets alarm priority level (1, 2

Seite 164

6-5BP1 Setup Factory DefaultsBP1 SETUP ADULT PED NEO DESCRIPTIONALARMS OFF OFF OFF Enables BP1 alarmALARM LEVEL 2 2 2 Sets alarm priority level (1, 2,

Seite 165

6-6BP2 Setup Factory DefaultsBP2 SETUP ADULT PED NEO DESCRIPTIONALARMS OFF OFF OFF Enables BP2 alarmALARM LEVEL 2 2 2 Sets alarm priority level (1, 2,

Seite 166

6-7BP3 Setup Factory DefaultsBP3 SETUP ADULT PED NEO DESCRIPTIONALARMS OFF OFF OFF Enables BP3 alarmALARM LEVEL 2 2 2 Sets alarm priority level (1, 2,

Seite 167

6-8NIBP Setup Factory DefaultsNIBP SETUP ADULT PED NEO DESCRIPTIONALARMS OFF OFF OFF Enables NIBP alarmALARM LEVEL 2 2 2 Sets alarm priority level (1,

Seite 168

6-9CO2 Setup Factory DefaultsCO2 SETUP ADULT PED NEO DESCRIPTIONALARMS OFF OFF OFF Enables CO2 alarmALARM LEVEL 2 2 2 Sets alarm priority level (1, 2,

Seite 169

6-10Temp1 Setup Factory DefaultsTEMP1 SETUP ADULT PED NEO DESCRIPTIONALARMS OFF OFF OFF Enables Temp1 alarmALARM LEVEL 2 2 2 Sets alarm priority level

Seite 170

7-1Chapter 7 – System & Parameter MessagesSystem and parameter messages remind, prompt, or warn you about the current condition of themonitor or i

Seite 171

7-2Battery MessagesMESSAGE BATTERY LEVELMONITOR BAT HI > 60% ChargeMONITOR BAT MID 20-60% ChargeMONITOR BAT LOW < 20% ChargeBATTERY VERY LOW(Int