12-8
POWER
Modular Batteries: 1-each 12V; 2.0 AH Rechargeable Sealed Lead Acid
Modular Battery Capacity: 35 charges @ 360 joules or 5 hours pacing at100 mA 70 ppm
Battery Charge Time: 6 hours to 90%
AC Main: 115/230 VAC selectable, 48 to 62 Hz autosensing
Power Consumption: Normal – 2 watts, charging battery – 12 watts, charging paddles, - 135 volts
DEFIBRILLATOR
ECG Source: Monitor leads or defib paddles, selectable
Waveform: Damped sinusoid per AAMI spec
Energy Selection: 2, 3, 5, 10, 20, 30, 50, 100, 150, 200, 300, 360 joules
Charge Time: To 360 joules, 12 seconds or less
Auto Internal Discharge: 60 seconds after charge, or power off
Armed Indication: Charge tone, indicator on front panel and paddle
Synchronizer: Sync discharge within 60 msec of R-wave peak
Paddles; Standard-Adult anterior/anterior; Optional-internal paddles, hands-off,
and pediatric paddles to slide over adult paddles
NONINVASIVE PACER
Type: Selectable, demand or asynchronous
Rate: 40, 60, 70, 80, 90, 100, 120, 140, 160 ppm +/- 3% or +/- 3 ppm
Waveform: Monophasic truncated exponential
Current: 10 to 80 mA, continuously adjustable
Leads Off Detect: Pace electrode lead fail indicator
CERTIFICATION
UL 2601-1, IEC 60601-1, IEC 60601-2
ENVIRONMENTAL
See shipping carton for storage and transport specifications