Battery Messages
MESSAGE BATTERY LEVEL
MONITOR BAT HI > 60% Charge
MONITOR BAT MID 20-60% Charge
MONITOR BAT LOW < 20% Charge
BATTERY VERY LOW
(Intermittent Alarm Tone Sounds)
Approximately 10 minutes of battery life remains
Parameter Messages
The following tables explain messages that are parameter specific. Messages concerning
parameters displayed in the WDZ will be displayed in a video box below the numeric data for the
corresponding waveform parameters. The message is displayed as long as the message condition
exists. All messages that apply to parameters in the numeric-only zone are displayed in the
numeric display area.
ECG/Respiration Messages
ECG/RESP MESSAGE MEANING
CVA Cardiovascular artifact (CVA) coincidence has been detected on
respiration. Respiration will reject coincident respiration while
the message is displayed. At least 8 out of 10 coincident
respirations are required to initiate and maintain the CVA
message.
APNEA The apnea alarm limits have been exceeded.
APN ALM OFF Respiration alarms are ON, but apnea alarm is OFF.
CHK LEADS One or more electrodes or lead wires may be dry or loose. In
addition to the message, the ECG waveform and respiration
waveform, if it is displayed, become dotted lines.