7-3
SPO2 Messages
SPO2 MESSAGE MEANING
SPO2 SRCH
(Nellcor only)
The monitor cannot locate the patient’s pulse. The patient’s
perfusion may be too poor to detect an acceptable pulse. Confirm
proper application of the sensor; make sure the ECG parameter is
available for C-LOCK synchronization; try another sensor site; or
try the OXISENSOR II R-15 sensor.
NO C-LOCK
(Nellcor only)
Indicates loss of synchronization between the ECG and SPO2
pulse waveform. Take steps to eliminate ECG artifact, or if the R-
wave amplitude is less than 4 mm, increase the ECG waveform
size by using the ECG SIZE softkey.
NO SENSOR The SPO2 parameter is ON, but the sensor and/or sensor cable
is/are not connected to the monitor.
SPO2 CAL The monitor is performing automatic self-calibration. The
calibration range is 50 to 100%.
Invasive Blood Pressure (BP1/BP2/BP3) Messages
BP1/BP2/BP3 MESSAGE MEANING
NO XDUCER The pressure transducer is not connected to the monitor.
ZERO BP1
ZERO BP2
ZERO BP3
Message flashes until the pressure channel is zeroed as a
reminder that pressures must be zeroed prior to use.
NO ZERO: PULSE DETECT Displayed over the BP SETUP page to indicate that either a pulse
or electrical noise was detected, which precluded zeroing the
transducer. Confirm that the transducer is vented to air before
pressing the ZERO key.
ZERO OUT OF RANGE Displayed over the BP SETUP page to indicate that a
transducer’s offset exceeds the zero range (±100 mmHg) of the
monitor. Confirm that the transducer is vented to the air and
attempt to zero again.