NOTE: Episodic measurements will not post to a VS chart more
than once for all automatic entries, even if they are displayed and
within the Valid Time. There is one exception to this rule: the
STORE VS softkey will capture ALL measurements displayed on the
monitor at a given time, regardless of their continuous or episodic
nature. The purpose of this exception is to ensure that ALL vital
signs displayed at the time the user selects the STORE VS softkey
will be sent to the VSC table and CIS.
Setting Vital Sign Storage Options
To set up storage options for vital sign charts, you must first either pop-up VSC or assign
VSC to a waveform trace. The following illustrates how to set up storage options while
VSC is in a pop-up:
1. Press the SYSTEM SETUP key to open the System Setup menu.
CLEAR
PT
DATA
STORE
VS
DISP
VSC
PT
SETUP
HR
AUTO
ECG
TRACE
SEL
MORE
2. Press the DISP VSC softkey to display VSC in the waveform display zone and to
display the following menu.
VSC
SETUP
STORE
VS
TABLE
→ →
TABLE
↑ ↑
TABLE
↓ ↓
TEXT
SIZE
MORE
3. Press the VSC SETUP softkey. The following VSC setup softkeys will be displayed.
4HR
OFF
1MIN
STORE
ALM
SETUP
STORE
NIBP
ON
STORE
TMP
ON
STORE
CO
OFF
4. On this screen you can:
Select time intervals between automatic VSC entries
Press the STORE ALM SETUP softkey to define which parameter alarms will
generate VSC entries:
(a) Toggle storage on all parameter alarms with STORE ALM ON softkey
(b) Touch the vital sign softkey on the menu to select a parameter
(c) Toggle storage on alarms for selected parameter with the ON/OFF softkey
(d) Select next parameter and on/off status, or MORE to exit menu
Turn storage on NIBP measurement ON/OFF
Turn storage on predictive TEMP readings ON/OFF
Turn storage on Cardiac Output measurement ON/OFF
NOTE: Options for parameters that are not installed will not
display on this screen.