6-10
Temp1 Setup Factory Defaults
TEMP1 SETUP ADULT PED NEO DESCRIPTION
ALARMS OFF OFF OFF Enables Temp1 alarm
ALARM LEVEL 2 2 2 Sets alarm priority level (1, 2, or 3)
LATCH ALARMS OFF OFF OFF Latches the Temp1 alarm until reset
UNITS
ο
C
ο
C
ο
C
Celsius or Fahrenheit
HIGH LIMIT 38.0 38.0 37.5 High alarm limit setting
LOW LIMIT 36.0 36.0 36.0 Low alarm limit setting
PRAM AVAIL ON ON ON Temp1 enable or disable
PRAM ON/OFF OFF OFF OFF Temp1 standby
TONE ON ON ON Measurement update tone enable
PRAM COLOR WHITE WHITE WHITE Sets on-screen color of parameter
Temp2 Setup Factory Defaults
TEMP2 SETUP ADULT PED NEO DESCRIPTION
ALARMS OFF OFF OFF Enable Temp2 alarm
ALARM LEVEL 2 2 2 Sets alarm priority level (1, 2, or 3)
LATCH ALARMS OFF OFF OFF Latches the Temp2 alarm until reset
HIGH LIMIT 38.0 38.0 37.5 High alarm limit setting
LOW LIMIT 36.0 36.0 36.0 Low alarm limit setting
PRAM AVAIL ON ON ON Temp2 enable or disable
PRAM ON/OFF OFF OFF OFF Temp2 standby
PRAM COLOR GRAY GRAY GRAY Sets on-screen color of parameter