3-24
1. Position nasal cannula on patient.
2. Connect ETCO2 tubing to Prism SE’s connector panel.
3. Observe WARM UP message on screen. ETCO2 monitoring will commence approximately
10 seconds following connection. If an error message appears, check tubing for blockage. If
this does not correct problem, see the Troubleshooting section.
Graphical Trends
Graphical trend allows view of compilation of data collected for given parameter over defined
time period. Time period choices are 1, 8, and 24 hours for each monitored parameter.
1. Press the SYSTEM SETUP key to open the System Setup menu.
CLEAR
PT
DATA
STORE
VS
DISP
VSC
PT
SETUP
HR
AUTO
ECG
TRACE
SEL
MORE
2. Press the MORE softkey twice to access the TREND softkey.
CONF TREND TEST
3. Press TREND softkey to display Trend Setup Page. Trend is displayed in TRACE 3.
4. Press the softkeys to alter display for parameter, trend time base, and scales. Use arrows to
move time visible on graph.
HR
RESP
SPO2
1 HR
8HR
24HR
SCALE
250
200
CURS
CURS
5. TREND can be recorded with use of bedside recorder. Press the RECORD key and within
two seconds press the softkey next to TRACE 3 (TREND GRAPH).
6. Press PAGE HOME to remove trend display and return to main monitoring
display.