SPO2 Sensor Accessories—Nellcor
Nellcor Sensors – Contact Nellcor
Sensor Model Patient size
Oxisensor® II oxygen transducers
(Sterile, Single-use only)
N-25/N-25LF
I-20
D-20
D-25/D25-L
R15
<3 or > 40Kg
3 to 20 Kg
10 to 50Kg
>30Kg
>50Kg
Oxiband oxygen transducer
(reusable with nonsterile adhesive)
OXI-A/N
OXI-P/I
<3 or > 40 Kg
<3 to 40Kg
Durasensor oxygen transducer
(Reusable, Nonsterile)
DS-100A >40Kg
OxiCliq oxygen transducers
(sterile, single use only)
P
N
I
A
10 to 50 Kg
<3 or > 40Kg
3 to 20 Kg
>30Kg
Nellcor MP405 SPO2 Ext Cable –
10 Foot, must be used with the
above sensors
E2800-64
WARNING: A pulse oximeter is an early warning device. Use a lab
co-oximeter to completely understand the patient’s condition.
NOTE: Inaccurate measurements or loss of pulse signal may be
caused by:
Incorrect sensor application or use
Significant levels of dysfunctional hemoglobins (HbCO or MetHb)
Intravascular dyes such as indocyanine green or methylene blue
Darkly pigmented skin
Exposure to excessive illumination, such as surgical lamps
(especially ones with a xenon light source), bilirubin lamps,
fluorescent lights, infrared heating lamps, or direct sunlight
(exposure to excessive illumination can be corrected by covering
sensor with a dark or opaque material)
Excessive patient movement
Venous pulsations
Placement of a sensor on an extremity with a blood pressure cuff,
arterial catheter, or intravascular line
NOTE: Loss of pulse signal can occur in any of the following
situations:
Sensor is too tight
Patient has hypotension, severe vasoconstriction, severe anemia,
or hypothermia